What’s Vulnerable › Vulnerabilities
CVE-2015-1012
- Severity: high (CVSS 7.5 v3.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.8% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2019-03-25 · updated 2026-06-17
- Weakness: CWE-312
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Affected products
- Pfizer Lifecare Pca Infusion System Firmware: through 5.0
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01