Description
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
Affected products
- Hospira Lifecare Pca3: all versions
- Hospira Lifecare Pca5: all versions
- Hospira Lifecare Pcainfusion Firmware: through 5.0
References
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- http://www.securityfocus.com/bid/75136
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01B
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- http://www.securityfocus.com/bid/75136
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01B