What’s Vulnerable › Vulnerabilities
CVE-2017-12716
- Severity: medium (CVSS 6.5 v3.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2018-04-25 · updated 2026-06-17
- Weakness: CWE-311
Description
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Affected products
- Abbott Accent Firmware: before f0b.0e.7e — fixed in f0b.0e.7e
- Abbott Accent Mri Firmware: before f10.08.6c — fixed in f10.08.6c
- Abbott Accent St Firmware: before f10.08.6c — fixed in f10.08.6c
- Abbott Anthem Firmware: before f0b.0e.7e — fixed in f0b.0e.7e
How to fix
Upgrade to a fixed version: Abbott Accent Firmware f0b.0e.7e; Abbott Accent Mri Firmware f10.08.6c; Abbott Accent St Firmware f10.08.6c; Abbott Anthem Firmware f0b.0e.7e. Check the vendor advisory for mitigations.
References
- http://www.securityfocus.com/bid/100523
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
- http://www.securityfocus.com/bid/100523
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01
Get alerted when CVE-2017-12716 or the next vulnerability affects your stack