What’s Vulnerable › Vulnerabilities
CVE-2018-14781: Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
- Severity: medium (CVSS 5.3 v3.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.7% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2018-08-13 · updated 2026-06-17
- Weakness: CWE-294
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Affected products
- Medtronicdiabetes 508 Minimed Insulin Pump Firmware: all versions
- Medtronicdiabetes 522 Paradigm Real-Time Firmware: all versions
- Medtronicdiabetes 523 Paradigm Revel Firmware: all versions
- Medtronicdiabetes 523K Paradigm Revel Firmware: all versions
- Medtronicdiabetes 551 Minimed 530G Firmware: all versions
- Medtronicdiabetes 722 Paradigm Real-Time Firmware: all versions
- Medtronicdiabetes 723 Paradigm Revel Firmware: all versions
- Medtronicdiabetes 723K Paradigm Revel Firmware: all versions
- Medtronicdiabetes 751 Minimed 530G Firmware: all versions
References
- http://www.securityfocus.com/bid/105044
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02
- http://www.securityfocus.com/bid/105044
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02
Get alerted when CVE-2018-14781 or the next vulnerability affects your stack