CVE-2018-8868: Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
- Severity: medium (CVSS 6.4 v3.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.3% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2018-07-03 · updated 2026-06-17
- Weakness: CWE-749
Description
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Affected products
- Medtronic 24950 Mycarelink Monitor Firmware: all versions
- Medtronic 24952 Mycarelink Monitor Firmware: all versions
References
- https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01