- Severity: critical (CVSS 10.0 v3.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 2.5% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2019-06-13 · updated 2026-06-17
- Weakness: CWE-434
Description
BD Alaris Gateway Workstation Versions, 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14, 1.3.1 Build 13, This does not impact the latest firmware Versions 1.3.2 and 1.6.1, Additionally, the following products using software Version 2.3.6 and below, Alaris GS, Alaris GH, Alaris CC, Alaris TIVA, The application does not restrict the upload of malicious files during a firmware update.
Affected products
- Bd Alaris Cc Syringe Pump Firmware: through 2.3.6
- BD Alaris Gateway Workstation: 1.1.3-10, 1.1.3-11, 1.2-15, 1.3.0-14, 1.3.1-13
- Bd Alaris Gh Syringe Pump Firmware: through 2.3.6
- Bd Alaris Gs Syringe Pump Firmware: through 2.3.6
- Bd Alaris Tiva Syringe Pump Firmware: through 2.3.6
References
- http://www.securityfocus.com/bid/108765
- https://ics-cert.us-cert.gov/advisories/ICSMA-19-164-01
- https://www.bd.com/en-us/support/product-security-and-privacy/product-security-bulletins/alaris-gateway-workstation-unauthorized-firmware
- http://www.securityfocus.com/bid/108765
- https://ics-cert.us-cert.gov/advisories/ICSMA-19-164-01