What’s Vulnerable › Vulnerabilities
CVE-2019-25720: Dräger SC Monitoring Devices DoS via Malformed Network Packet
- Severity: medium (CVSS 6.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2026-06-03 · updated 2026-07-22
- Weakness: CWE-1286
Description
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Affected products
- Dräger SC6802XL: SC6802XL
- Dräger SC8000: SC8000
- Dräger SC90000 XL: SC90000 XL
- Dräger SC 6002XL: SC 6002XL
- Dräger SC 7000: SC 7000
References
- https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf
- https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet