What’s Vulnerable › Vulnerabilities
CVE-2020-12035
- Severity: medium (CVSS 4.9 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2020-06-29 · updated 2026-06-17
- Weakness: CWE-287
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
Affected products
- Baxter Prismaflex Firmware: all versions
- Baxter Prismax Firmware: before 3.0 — fixed in 3.0
How to fix
Upgrade to a fixed version: Baxter Prismax Firmware 3.0. Check the vendor advisory for mitigations.
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01
- https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01
Get alerted when CVE-2020-12035 or the next vulnerability affects your stack