What’s Vulnerable › Vulnerabilities
CVE-2020-12036
- Severity: high (CVSS 7.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.5% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2020-06-29 · updated 2026-06-17
- Weakness: CWE-319
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Affected products
- Baxter Prismaflex Firmware: all versions
- Baxter Prismax Firmware: before 3.0 — fixed in 3.0
How to fix
Upgrade to a fixed version: Baxter Prismax Firmware 3.0. Check the vendor advisory for mitigations.
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01