What’s Vulnerable › Vulnerabilities
CVE-2020-12048
- Severity: high (CVSS 7.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.5% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2020-06-29 · updated 2026-06-17
- Weakness: CWE-319
Description
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Affected products
- Baxter Phoenix Hemodialysis Delivery System: 3.36, 3.40
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-03
- https://www.us-cert.gov/ics/advisories/icsma-20-170-03
Get alerted when CVE-2020-12048 or the next vulnerability affects your stack