CVE-2020-12495: ENDRESS+HAUSER: Ecograph T utilizing Webserver firmware version 1.x has improper privilege management
- Severity: high (CVSS 8.8 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.9% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2020-11-19 · updated 2026-06-17
- Weakness: CWE-269
Description
Endress+Hauser Ecograph T (Neutral/Private Label) (RSG35, ORSG35) with Firmware version prior to V2.0.0 is prone to improper privilege management. The affected device has a web-based user interface with a role-based access system. Users with different roles have different write and read privileges. The access system is based on dynamic "tokens". The vulnerability is that user sessions are not closed correctly and a user with fewer rights is assigned the higher rights when he logs on.
Affected products
- Endress Orsg35 Firmware: before 2.0.0 — fixed in 2.0.0
- Endress Orsg45 Firmware: before 2.0.0 — fixed in 2.0.0
- Endress Rsg35 Firmware: before 2.0.0 — fixed in 2.0.0
- Endress Rsg45 Firmware: before 2.0.0 — fixed in 2.0.0
How to fix
Upgrade to a fixed version: Endress Orsg35 Firmware 2.0.0; Endress Orsg45 Firmware 2.0.0; Endress Rsg35 Firmware 2.0.0; Endress Rsg45 Firmware 2.0.0. Check the vendor advisory for mitigations.
References
- https://cert.vde.com/en-us/advisories/vde-2020-021
- https://cert.vde.com/en-us/advisories/vde-2020-021
Get alerted when CVE-2020-12495 or the next vulnerability affects your stack