CVE-2020-6963
- Severity: critical (CVSS 10.0 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 2.7% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2020-01-24 · updated 2026-06-17
- Weakness: CWE-798
Description
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
Affected products
- Gehealthcare Apexpro Telemetry Server Firmware: through 4.2
- Gehealthcare Carescape Central Station Mai700 Firmware: 1.0
- Gehealthcare Carescape Central Station Mas700 Firmware: 1.0
- Gehealthcare Carescape Telemetry Server Mp100R Firmware: through 4.2
- Gehealthcare Clinical Information Center Mp100D Firmware: 4.0, 5.0
- Gehealthcare Clinical Information Center Mp100R Firmware: 4.0, 5.0
References
- https://www.us-cert.gov/ics/advisories/icsma-20-023-01
- https://www3.gehealthcare.com/~/media/downloads/us/support/site-planning/site-readiness/gehc-gateway_project_implementation_guide_pdf.pdf
- https://www.us-cert.gov/ics/advisories/icsma-20-023-01