What’s Vulnerable › Vulnerabilities
CVE-2021-33882
- Severity: high (CVSS 8.6 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 1.1% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2021-08-25 · updated 2026-06-17
- Weakness: CWE-306
Description
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.
Affected products
- Bbraun Spacecom2: before 012u000062 — fixed in 012u000062
How to fix
Upgrade to a fixed version: Bbraun Spacecom2 012u000062. Check the vendor advisory for mitigations.
References
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Get alerted when CVE-2021-33882 or the next vulnerability affects your stack