What’s Vulnerable › Vulnerabilities
CVE-2021-33885
- Severity: critical (CVSS 9.8 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 5.6% probability of exploitation in the next 30 days
- Public exploit or detection template: yes
- Published: 2021-08-25 · updated 2026-06-17
- Weakness: CWE-347
Description
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
Affected products
- Bbraun Spacecom2: before 012u000062 — fixed in 012u000062
How to fix
Upgrade to a fixed version: Bbraun Spacecom2 012u000062. Check the vendor advisory for mitigations.
References
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
- https://www.bbraunusa.com/en.htm
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/