What’s Vulnerable › Vulnerabilities
CVE-2021-41835: Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
- Severity: high (CVSS 7.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.3% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2022-01-21 · updated 2026-06-17
- Weakness: CWE-327
Description
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port TCP/80 to the encrypted service.
Affected products
- Fresenius-Kabi Agilia Connect: through d25
- Fresenius-Kabi Agilia Partner Maintenance Software: through 3.3.0
- Fresenius-Kabi Link+ Agilia Firmware: before 3.0; 3.0, 3.0-d15 — fixed in 3.0
- Fresenius-Kabi Vigilant Centerium: 1.0
- Fresenius-Kabi Vigilant Insight: 1.0
- Fresenius-Kabi Vigilant Mastermed: 1.0
How to fix
Upgrade to a fixed version: Fresenius-Kabi Link+ Agilia Firmware 3.0. Check the vendor advisory for mitigations.
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01
- https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01