What’s Vulnerable › Vulnerabilities
CVE-2022-3027: Contec Health CMS8000
- Severity: medium (CVSS 5.7 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.3% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2022-09-13 · updated 2026-06-17
- Weakness: CWE-284
Description
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Affected products
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01
Get alerted when CVE-2022-3027 or the next vulnerability affects your stack