What’s Vulnerable › Vulnerabilities
CVE-2022-38100: Contec Health CMS8000
- Severity: high (CVSS 7.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 1.0% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2022-09-13 · updated 2026-06-17
- Weakness: CWE-400
Description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Affected products
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01