What’s Vulnerable › Vulnerabilities
CVE-2022-4992: Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering
- Severity: high (CVSS 8.6 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2026-06-02 · updated 2026-07-22
- Weakness: CWE-345
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
Affected products
- Dräger Infinity Acute Care System: before VG4.2; before VG4.1.1; before VG4.0.3 — fixed in VG4.0.3, VG4.1.1, VG4.2
- Dräger Standalone Infinity M540 patient monitor: before VG4.2; before VG4.1.1 — fixed in VG4.1.1, VG4.2
How to fix
Upgrade to a fixed version: Dräger Infinity Acute Care System VG4.0.3, VG4.1.1, VG4.2; Dräger Standalone Infinity M540 patient monitor VG4.1.1, VG4.2. Check the vendor advisory for mitigations.
References
- https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf
- https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering
Get alerted when CVE-2022-4992 or the next vulnerability affects your stack