What’s Vulnerable › Vulnerabilities
CVE-2023-0888: Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
- Severity: high (CVSS 7.2 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.6% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2023-03-13 · updated 2026-06-17
- Weakness: CWE-95
Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Affected products
- B. Braun Space Battery Pack SP with Wi-Fi: through 053l000092; through 054u000092
References
- https://www.bbraun.com/productsecurity
- https://www.bbraunusa.com/productsecurity
- https://www.bbraun.com/productsecurity
- https://www.bbraunusa.com/productsecurity
Get alerted when CVE-2023-0888 or the next vulnerability affects your stack