CVE-2024-48973: Debug port on Life2000 Ventilator serial interface is enabled by default
- Severity: critical (CVSS 9.3 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2024-11-14 · updated 2026-06-17
- Weakness: CWE-1263
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Affected products
- Baxter Life2000 Ventilator Firmware: through 06.08.00.00