What’s Vulnerable › Vulnerabilities
CVE-2024-48974: Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
- Severity: critical (CVSS 9.3 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2024-11-14 · updated 2026-06-17
- Weakness: CWE-494
Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Affected products
- Baxter Life2000 Ventilator Firmware: through 06.08.00.00
References