What’s Vulnerable › Vulnerabilities
CVE-2024-9832: No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
- Severity: critical (CVSS 9.3 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.2% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2024-11-14 · updated 2026-06-17
- Weakness: CWE-307
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected products
- Baxter Life2000 Ventilator Firmware: through 06.08.00.00
References
Get alerted when CVE-2024-9832 or the next vulnerability affects your stack