What’s Vulnerable › Vulnerabilities
CVE-2025-0626: Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
- Severity: high (CVSS 7.5 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 1.1% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-01-30 · updated 2026-06-17
- Weakness: CWE-912
Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Affected products
- Contec Health CMS8000 Patient Monitor: all versions
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
- https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
Get alerted when CVE-2025-0626 or the next vulnerability affects your stack