What’s Vulnerable › Vulnerabilities
CVE-2025-0683: Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
- Severity: medium (CVSS 5.9 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.8% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-01-30 · updated 2026-06-17
- Weakness: CWE-359
Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Affected products
- Contec Health CMS8000 Patient Monitor: all versions
References
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
- https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
Get alerted when CVE-2025-0683 or the next vulnerability affects your stack