What’s Vulnerable › Vulnerabilities
CVE-2025-1204
- Severity: high (CVSS 7.7 v4.0)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.5% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-02-25 · updated 2026-06-17
- Weakness: CWE-912
Description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Affected products
References
- https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
Get alerted when CVE-2025-1204 or the next vulnerability affects your stack