What’s Vulnerable › Vulnerabilities
CVE-2025-14096: Credential Disclosure vulnerability in Radiometer Products
- Severity: high (CVSS 8.4 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.1% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-12-17 · updated 2026-06-17
- Weakness: CWE-250
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Affected products
- Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers: Windows 7 Operating system, Windows XP Operating system
- Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers: Windows 7 Operating system, Windows XP Operating system
- Radiometer Medical Aps AQT90 FLEX Analyzers: Windows 7 Operating system, Windows XP Operating system
References
Get alerted when CVE-2025-14096 or the next vulnerability affects your stack