What’s Vulnerable › Vulnerabilities
CVE-2025-53963
- Severity: critical (CVSS 9.8 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.5% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-12-04 · updated 2026-06-17
- Weakness: CWE-521
Description
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
Affected products
- Thermofisher Ion Torrent Onetouch 2 Firmware: all versions
References
- https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf
- https://tools.thermofisher.cn/content/sfs/brochures/One_Touch_2_Spec_Sheet.pdf
Get alerted when CVE-2025-53963 or the next vulnerability affects your stack