What’s Vulnerable › Vulnerabilities
CVE-2025-64778: Mirion Medical EC2 Software NMIS BioDose Use of Hard-coded Credentials
- Severity: high (CVSS 7.8 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.1% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2025-12-02 · updated 2026-09-25
- Weakness: CWE-798
Description
NMIS/BioDose software V22.02 and previous versions contain executable binaries with plain text hard-coded passwords. These hard-coded passwords could allow unauthorized access to both the application and database.
Affected products
- Mirion Biodose/Nmis: before 23.0 — fixed in 23.0
How to fix
Upgrade to a fixed version: Mirion Biodose/Nmis 23.0. Check the vendor advisory for mitigations.
References
Get alerted when CVE-2025-64778 or the next vulnerability affects your stack