What’s Vulnerable › Vulnerabilities
CVE-2026-18164: Flow Neuroscience FL-100 Use of Hard-coded Credentials
- Severity: high (CVSS 8.1 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.4% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2026-08-13 · updated 2026-09-08
- Weakness: CWE-798
Description
An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarily manipulate brain stimulation parameters and state.
Affected products
- Flow Neuroscience FL-100: before July 2026 — fixed in July 2026
- Halo Neuroscience FL-100: before July 2026 — fixed in July 2026
How to fix
Upgrade to a fixed version: Flow Neuroscience FL-100 July 2026; Halo Neuroscience FL-100 July 2026. Check the vendor advisory for mitigations.
References
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-225-01.json
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-225-01
Get alerted when CVE-2026-18164 or the next vulnerability affects your stack