What’s Vulnerable › Vulnerabilities
CVE-2026-64934: Mira Hormone Monitor, Mira Android App Reliance on untrusted inputs in a security decision
- Severity: medium (CVSS 4.3 v3.1)
- Exploited in the wild: not listed in CISA KEV
- EPSS: 0.3% probability of exploitation in the next 30 days
- Public exploit or detection template: none known
- Published: 2026-08-11 · updated 2026-09-01
- Weakness: CWE-807
Description
The Mira cloud API accepts the firmware version reported by the companion app as authoritative for a given device, without independently attesting the version from the device itself. An authenticated attacker could submit arbitrary firmware version strings for their own device, allowing them to evade vendor-side vulnerable-fleet analytics, suppress security update prompts to the user, and misrepresent patch-adoption metrics.
Affected products
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Android App: 4.5.15.4
- Quanovate Tech Inc. (operating as Mira / Mira Care) Mira Firmware: 1.7.1.47
References
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01
Get alerted when CVE-2026-64934 or the next vulnerability affects your stack